Dose analysis of the Model 112A Pulserad Pulsed X-Ray Generator by ITS/CYLTRAN.

Loading...
Thumbnail Image
Authors
Kim, Jong Ryul
Subjects
flash x-ray
ITS/CYLTRAN
Monte Carlo
Advisors
Maruyama, Xavier K.
Date of Issue
1989-12
Date
December 1989
Publisher
Monterey, California. Naval Postgraduate School
Language
en_US
Abstract
The radiation output from the NPS Pulserad 112A X—ray Generator has been calculated using the ITS code CYLTRAN. The results of this numerical simulation have been compared to the experimental measurements taken previously. These calculations show that the overall photon energy spectrum is independent of the electron source beam parameters such as beam radius and angular distribution. A previously unexplained measured dip in the radiation dose at the beam center line can be explained with an angular divergence of the electron beam from the cathode. The simulation is successful in explaining the pattern of the dose distribution, but the calculation is in general much smaller than the measured values. This inability to reproduce the magnitude of the dose pattern points out the need to measure the radiation dose each time the Pulserad 112A is used. Some suggestions for explaining this discrepancy are included.
Type
Thesis
Description
Series/Report No
Department
Physics
Organization
Naval Postgraduate School (U.S.)
Identifiers
NPS Report Number
Sponsors
Funder
Format
74 p.
Citation
Distribution Statement
Approved for public release; distribution is unlimited.