Studies and analyses in support of the oil analysis program

Loading...
Thumbnail Image
Authors
Jayachandran, Toke
Larson, Harold J.
Subjects
spectrometer
correlation program
calibration atomic emission, atomic absorption
Advisors
Date of Issue
1986-01
Date
Project Report for Period October 1984 - December 1985
Publisher
Monterey, California. Naval Postgraduate School
Language
en_US
Abstract
Results of statistical analyses of the data from the field test of a newly developed Portable Wear Metal Analyzer (PWMA) are reported. New formulas for converting readings from an atomic absorption spectrometer into equivalent readings for an atomic emission spectrometer are derived. Efforts to provide technical consultation on the development of a new software package to computerize some of the daily routine operations of oil laboratories are discussed.
Type
Technical Report
Description
Series/Report No
Department
Organization
Graduate School of Operational and Information Sciences (GSOIS)
Identifiers
NPS Report Number
NPS-53-86-0003PR
Sponsors
Funder
Format
Citation
Distribution Statement
Rights