Show simple item record

dc.contributor.authorRothauge, Charles H.
dc.contributor.authorHamburger, Ferdinand Jr.
dc.date.accessioned2017-06-29T16:59:09Z
dc.date.available2017-06-29T16:59:09Z
dc.date.issued1950-10
dc.identifier.citationC.H. Rothauge, F. Hamburger, Jr., "Measurement of the electrical characteristics of quartz crystal units by use of a bridged-tee network," Proceedings of the I.R.E., (October 1950), pp.1213-1216.en_US
dc.identifier.urihttp://hdl.handle.net/10945/55135
dc.description.abstractA bridged-tee null network has been used to measure the equivalent resistance and the equivalent reactance of a quartz crystal plate. This measuring circuit has the advantages that shielding is relatively simple (both the source and the detector have a common grounded terminal) and that corrections for all stray capacitances that affect the measurements may be included in the calibrations of the capacitors of the tee. The precision of the measurements is estimated to be 0.3 per cent for the determination of the equivalent reactance and 2.3 per cent for the determination of the equivalent resistance.en_US
dc.format.extent4 p.en_US
dc.publisherInstitute of Radio Engineersen_US
dc.rightsThis publication is a work of the U.S. Government as defined in Title 17, United States Code, Section 101. Copyright protection is not available for this work in the United States.en_US
dc.titleMeasurement of the electrical characteristics of quartz crystal units by use of a bridged-tee networken_US
dc.typeArticleen_US
dc.contributor.corporateNaval Postgraduate School (U.S.)en_US
dc.contributor.departmentElectrical Engineeringen_US


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record